Section II: Object
Capillary eletrophoresis instrument
The Department of Forensic Medicine, Section of Forensic Genetics.
Fluorescent-marked PCR products from the DNA profile analysis in crime cases are separated and
detected on Capillary Electrophoresis (CE) instruments.
luorescent-marked PCR products from the DNA profile analysis in crime cases are separated and detected on Capillary Electrophoresis (CE) instruments. The instrument must be able to:
1) separate minimum 6 fluorescent colors
2) separate PCR fragments with length differences < 1 bp
3) 80 – 500 bp
4) have at least 24 capillaries
5) Electrokinetic injection
6) < 1 hour/injection
To our knowledge, there are no other instruments than Thermo Fishers AB3500XL and SeqStudio Flex on the marked, that meet these requirements.
Besides, AB3500XL and SeqStudioFlex are already tested and validated at DFI which makes
implementation/validation of new instrument shorter and low-cost.
The value of this procurement is below the threshold value.
Section IV: Procedure
- The works, supplies or services can be provided only by a particular economic operator for the following reason:
- absence of competition for technical reasons
- Additional deliveries by the original supplier ordered under the strict conditions stated in the directive
Fluorescent-marked PCR products from the DNA-profile analysis in crime cases are separated and detected on Capillary
Electrophoresis (CE) instruments. The instrument must be able to:
1) separate minimum 6 fluorescent colors
2) separate PCR fragments with length differences < 1 bp
3) 80 – 500 bp
4) have at least 24 capillaries
5) Electrokinetic injection J. nr. 036-0010/18-3600 4
6) < 1 hour/injection
To our knowledge, there are no other instruments than Thermo Fishers AB3500XL and SeqStudio Flex on the marked, that meet these requirements.
Besides, AB3500XL and SeqStudioFlex are already tested and validated at DFI which makes implementation/validation of new
instrument shorter and low-cost.
There is already a SeqStudioFlex which has been tested and validated for use in crime cases. We cannot run samples on two different instruments due to downstream analysis of the results. An instrument from another vendor will always give different quantitative results. This will make a second validation and second analysis pipeline necessary.
All methods and analysis in crime cases are validated according to ISO 17025.
Section V: Award of contract/concession
National registration number: 24246299
Postal address: Kamstrupvej 90
Town: Alle
NUTS code: DK02 Sjælland
Postal code: 4000
Country: Denmark
Section VI: Complementary information
Please notice that the value of this procurement is under the threshold value.
Postal address: Nævnenes Hus, Toldboden 2
Town: Viborg
Postal code: 8800
Country: Denmark
E-mail: klfu@naevneneshus.dk
Telephone: +45 72405600
Internet address: www.KLFU.DK
Postal address: Carl Jacobsens Vej 35
Town: Valby
Postal code: 2500
Country: Denmark
E-mail: kfst@kfst.dk
Telephone: +45 41715000
Internet address: www.kfst.dk