Triple quadrupole ICP-MS/MS
Fødevarestyrelsen
Contract award notice
Results of the procurement procedure
Supplies
Section I: Contracting authority
62534516
Stationsparken 31-33
Glostrup
2600
Denmark
Contact person: Inge Rokkjær
Telephone: +45 72275476
E-mail: inro@fvst.dk
NUTS code: DK022
Internet address(es):
Main address: http://www.fvst.dk
Section II: Object
Triple quadrupole ICP-MS/MS.
The inductive coupled mass spectrometer (ICP-MS/MS) shall be with a broad elemental coverage, a highmatrix tolerance and low detection limits. The ICP-MS/MS shall also be able to support speciation analysis with chromatography coupling and provide a sensitive technique for monitoring nanoparticles such as SiO2, TiO2 and ZnO from food additives and packaging.
The ICP-MS/MS must be delivered in Lystrup, Denmark.
System Requirements:
Point 1:
The system shall be able to operate in MS/MS mode as stated in the IUPAC definition (IUPAC 2013Recommendations, Term 538) for a maximum of interference removal capacity for the more difficult matrix.
Point 2:
The instrument shall be able to analyse samples containing up to 25 % total dissolved solids (TDS) using thestandard sample introduction system by direct injection (no liquid dilution requested before) without switching the system interface.
Point 3:
The standard interface of the system shall be able to analyse samples with 25%TDS and offer a minimum sensitivity of 150 000 CPS/ppb for 7 Li, 500 000 CPS/ppb for 89 Y and 300 000 CPS/ppb for 205 Tl, and having a background of 0.2 CPS at mass 9 and 238 amu.
Point 4:
The system shall be capable of detecting and offering high abundance sensitivity (> 10-10) to avoid the impact of high concentrated element (at the mass M) on the mass at M+1 or M-1 amu, example: This feature allows the analysis of trace levels of manganese (55 Mn) in the presence of 1 000 ppm iron (56 Fe).
Point 5:
The detector shall be able to provide ultra-fast data acquisition, with 10 000 separate measurement per secondfor time resolved analysis (TRA) measurement of transient signals, which is required for the detection of nanoparticle under the Single Particle mode.
Point 6:
The electron multiplier detector must offer 10 orders of dynamic range for the analysis of trace and major elements in 1 sample measurement.
Point 7:
Discrete sampling system: It must be equipped with a valve to ensure automatic injection of the tuning solution without user interference for automatic optimatization of the ICP when tuning on the plasma. Discrete sampling system should be developed by the ICP provider to ensure the compatibility of the system with the software, the support on the items and the availability of the spare parts.
Point 8:
The system software: The software shall be able to handle direct control of our HPLC module (1200 systemfrom Agilent) so that only 1 software controls all the instruments when doing speciation analysis.
Point 9:
General requirement: All parts of the system must be delivered, installed and serviced by the supplier of the ICP-MS system.
Section IV: Procedure
- The works, supplies or services can be provided only by a particular economic operator for the following reason:
- absence of competition for technical reasons
Only one supplier (Agilent) can fulfil the criteria mentioned above in point 2-9, section II.2.4.
Section V: Award of contract
Triple quadrupole ICP-MS/MS
21852902
Produktionsvej 42
Glostrup
2600
Denmark
E-mail: customercare_denmark@agilent.com
NUTS code: DK022
Internet address:http://www.agilent.com
Section VI: Complementary information
Toldboden 2
Viborg
8800
Denmark
Telephone: +45 35291095
E-mail: klfu@erst.dk
Internet address:https://erhvervsstyrelsen.dk/klagenaevnet-for-udbud
Carl Jacobsensvej 35
Valby
2500
Denmark
Telephone: +45 41715000
E-mail: kfst@kfst.dk
Internet address:www.kfst.dk