Fakta om udbudet
Udbyder
Kungliga Tekniska högskolan
Vindere
Inverted Lattice Light Sheet Microscope
(24.05.2019)
Carl Zeiss AB, Division of Microscopy
PO Box 27324
102 54 Stockholm
Opdateringer
II.1.4)
Placing the text to be modified:Short description:
instead of:
For our research on subcellular signalling, organelle dynamics and intercellular communication we need to purchase an inverted lattice light sheet microscope that is capable to do fast and long term 4D fluorescence imaging of living cells.
Read:
For our research on subcellular signalling, organelle dynamics and intercellular communication we need to purchase a light sheet microscope that is capable to do fast and long term 4D fluorescence imaging of living cells.
II.2.4)
Placing the text to be modified:Description of the procurement:
instead of:
The inverted lattice light sheet microscope must operate with low light level illumination for minimal phototoxicity, have a lateral resolution of at least 300 nm, a field of view of at least 300 μm, have a volume (300 x 50 x 20 μmjavax.xml.bind.JAXBElement@7a61df2f) scan speed of at least 2,5 volumes/second and be compatible with standard sample carriers (e.g. 35 mm glass-bottom petri dishes).
Read:
The light sheet microscope must operate with low light level illumination for minimal phototoxicity.
IV.1.1)
Placing the text to be modified:Type of procedure
instead of:
Read:
Unstated
IV.1.1)
Placing the text to be modified:Type of procedure
instead of:
The inverted lattice light sheet microscope is a novel system protected by exclusive rights. It has unique functions (see below). Carls Zeiss is the sole producer of the equipment and Carl Zeiss AB the only distributor for sales to Sweden.
The inverted lattice light sheet microscope have the following functions and minimal specifications:
— complete inverted lattice light sheet microscope system optimized for live cell imaging,
— sample and optics need to be physically separated (coverslip) to avoid sample/optics contamination,
— samples prepared on standard coverslips #1,5,
— system compatible with standard sample carriers slides/coverslips, glass-bottom culture dishes (e.g. 35 mm), chamber slides, multiwell plates (6, 12, 24, …, 96, 384, 1096 wells),
— sample positioning and alignment fully motorized,
— motorized dispenser of immersion media for long time lapse recordings,
— environment control (C02, humidity and temperature) of sample chamber. Ports for perfusion in sample chamber,
— volume speed at least 2,5 Vol/s @ 300 μm x 50 μm x 20 μm,
— single plane speed at least 400 fps @ 300 μm x 20 μm,
— equipped with lasers for 488, 561, and 640 nm illumination and fast sequential multi-color imaging,
— fast switching of illumination beams (colors) within μ-seconds.,
— selection of specific light sheets via software (no direct physical interaction with hardware for alignment),
— optical system adjustment fully motorized and SW supported, no direct user interaction with hardware for adjustment required,
— control and recording computer system that can sustain a data rate permitting full speed recording for 1 h and time lapse imaging of 2 volumes/second for up to 24 hrs.
Read:
The light sheet microscope is a novel system protected by exclusive rights. It has unique functions. Carls Zeiss is the sole producer of the equipment and Carl Zeiss AB the only distributor for sales to Sweden.
Additional information
Visma notice: https://opic.com/id/afikwqpady
Inverted Lattice Light Sheet Microscope
Kungliga Tekniska högskolan
Voluntary ex ante transparency notice
Services
Directive 2014/24/EU
Section I: Contracting authority/entity
202100-3054
Brinellvägen 8
Stockholm
100 44
Sweden
Contact person: Karin Edoff
E-mail: kedoff@kth.se
NUTS code: SE11
Internet address(es):
Main address: http://www.kth.se
Section II: Object
Inverted Lattice Light Sheet Microscope
For our research on subcellular signalling, organelle dynamics and intercellular communication we need to purchase an inverted lattice light sheet microscope that is capable to do fast and long term 4D fluorescence imaging of living cells.
The inverted lattice light sheet microscope must operate with low light level illumination for minimal phototoxicity, have a lateral resolution of at least 300 nm, a field of view of at least 300 μm, have a volume (300 x 50 x 20 μm3) scan speed of at least 2,5 volumes/second and be compatible with standard sample carriers (e.g. 35 mm glass-bottom petri dishes).
Section IV: Procedure
- The works, supplies or services can be provided only by a particular economic operator for the following reason:
- protection of exclusive rights, including intellectual property rights
The inverted lattice light sheet microscope is a novel system protected by exclusive rights. It has unique functions (see below). Carls Zeiss is the sole producer of the equipment and Carl Zeiss AB the only distributor for sales to Sweden.
The inverted lattice light sheet microscope have the following functions and minimal specifications:
— complete Inverted Lattice Light sheet microscope system optimized for live cell imaging,
— sample and optics need to be physically separated (coverslip) to avoid sample/optics contamination,
— samples prepared on standard coverslips #1.5,
— system compatible with standard sample carriers slides/coverslips, glass-bottom culture dishes (e.g. 35 mm), chamber slides, multiwell plates (6, 12, 24, …, 96, 384, 1096 wells),
— sample positioning and alignment fully motorized,
— motorized dispenser of immersion media for long time lapse recordings,
— environment control (C02, humidity and temperature) of sample chamber. Ports for perfusion in sample chamber,
— volume speed at least 2,5 Vol/s @ 300 μm x 50 μm x 20 μm,
— single plane speed at least 400 fps @ 300 μm x 20 μm,
— equipped with lasers for 488, 561, and 640 nm illumination and fast sequential multi-color imaging,
— fast switching of illumination beams (colors) within μ-seconds,
— selection of specific light sheets via software (no direct physical interaction with hardware for alignment),
— optical system adjustment fully motorized and SW supported, no direct user interaction with hardware for adjustment required,
— control and recording computer system that can sustain a data rate permitting full speed recording for 1 h and time lapse imaging of 2 volumes/second for up to 24 hrs.
Section V: Award of contract/concession
Inverted Lattice Light Sheet Microscope
556035-3608
PO Box 27324
Stockholm
102 54
Sweden
Telephone: +46 84592500
E-mail: info.se@zeiss.com
NUTS code: SE110